Get Started Free →EU MDR 2017/745 compliance specialist for medical device classification, technical documentation, clinical evidence, and post-market surveillance. Covers Annex VIII classification rules, Annex II/III technical files, Annex XIV clinical evaluation, Art. 86 PSUR schedules, and EUDAMED integration. Use when classifying a medical device under MDR, building or gap-checking a technical file, planning clinical evaluation or PMS/PSUR cadence, or preparing for notified body review (e.g., 'what class is m
| Test case | Without → With | Effect | Δ tokens | Δ turns |
|---|---|---|---|---|
| case-17 | ✗→✓ | ▲ Improved | 133% | 0% |
| case-02 | ✓→✓ | = Same ✓ | 82% | 0% |
| case-01 | ✓→✓ | = Same ✓ | 125% | 0% |
| case-03 | ✓→✓ | = Same ✓ | 70% | 0% |
| case-04 | ✓→✓ | = Same ✓ | 181% | 0% |
EU MDR compliance patterns for medical device classification, technical documentation, and clinical evidence.
Classify device under MDR Annex VIII:
| Factor | Class I | Class IIa | Class IIb | Class III | |--------|---------|-----------|-----------|-----------| | Duration | Any | Short-term | Long-term | Long-term | | Invasiveness | Non-invasive | Body orifice | Surgical | Implantable | | System | Any | Non-critical | Critical organs | CNS/cardiac | | Risk | Lowest | Low-medium | Medium-high | Highest |
| Information Use | Condition Severity | Class | |-----------------|-------------------|-------| | Informs decision | Non-serious | IIa | | Informs decision | Serious | IIb | | Drives/treats | Critical | III |
Example 1: Absorbable Surgical Suture
Example 2: AI Diagnostic Software
Example 3: Cardiac Pacemaker
Prepare technical file per Annex II and III:
ANNEX II TECHNICAL DOCUMENTATION
├── Device description and UDI-DI
├── Label and instructions for use
├── Design and manufacturing info
├── GSPR compliance matrix
├── Benefit-risk analysis
├── Verification and validation
└── Clinical evaluation report| Requirement | Evidence | Status | |-------------|----------|--------| | Safe design (GSPR 1-3) | Risk management file | ☐ | | Chemical properties (GSPR 10.1) | Biocompatibility report | ☐ | | Infection risk (GSPR 10.2) | Sterilization validation | ☐ | | Software requirements (GSPR 17) | IEC 62304 documentation | ☐ | | Labeling (GSPR 23) | Label artwork, IFU | ☐ |
| Class | Route | NB Involvement | |-------|-------|----------------| | I | Annex II self-declaration | None | | Is/Im | Annex II + IX/XI | Sterile/measuring aspects | | IIa | Annex II + IX or XI | Product or QMS | | IIb | Annex IX, or Annex X + XI | QMS + tech doc assessment, or type exam + production | | III | Annex IX, or Annex X + XI | Full QMS + product dossier, or type exam + production |
Develop clinical evidence strategy per Annex XIV:
| Class | Minimum Evidence | Investigation | |-------|------------------|---------------| | I | Risk-benefit analysis | Not typically required | | IIa | Literature + post-market | May be required | | IIb | Systematic literature review | Often required | | III | Comprehensive clinical data | Required (Article 61) |
CER CONTENTS
├── Executive summary
├── Device scope and intended purpose
├── Clinical background (state of the art)
├── Literature search methodology
├── Data appraisal and analysis
├── Safety and performance conclusions
├── Benefit-risk determination
└── PMCF plan summaryEstablish PMS system per Chapter VII:
| Component | Requirement | Frequency | |-----------|-------------|-----------| | PMS Plan | Article 84 | Maintain current | | PSUR | Article 86 — Class IIa and higher | Per Art. 86(1) schedule below | | PMCF Plan | Annex XIV Part B | Update with CER | | PMCF Report | Annex XIV Part B | Annual (Class III) | | Vigilance | Articles 87-92 | As events occur |
| Class | Frequency (MDR Art. 86(1)) | |-------|-----------------------------| | Class III | Updated at least annually | | Class IIb (all, incl. implantable) | Updated at least annually | | Class IIa | When necessary, at least every 2 years | | Class I | No PSUR — PMS report instead (Art. 85) |
| Timeline | Requirement | |----------|-------------| | 2 days | Serious public health threat | | 10 days | Death or serious deterioration | | 15 days | Other serious incidents |
Implement UDI system per Article 27:
| Module | Content | Actor | |--------|---------|-------| | Actor | Company registration | Manufacturer, AR | | UDI/Device | Device and variant data | Manufacturer | | Certificates | NB certificates | Notified Body | | Clinical Investigation | Study registration | Sponsor | | Vigilance | Incident reports | Manufacturer | | Market Surveillance | Authority actions | Competent Authority |
Required elements per Article 13:
references/mdr-classification-guide.md contains:
references/clinical-evidence-requirements.md contains:
references/technical-documentation-templates.md contains:
bash# Quick gap analysis python scripts/mdr_gap_analyzer.py --device "Device Name" --class IIa # JSON output for integration python scripts/mdr_gap_analyzer.py --device "Device Name" --class III --output json # Interactive assessment python scripts/mdr_gap_analyzer.py --interactive
Analyzes device against MDR requirements, identifies compliance gaps, generates prioritized recommendations.
Output includes:
| Factor | Considerations | |--------|----------------| | Designation scope | Covers your device type | | Capacity | Timeline for initial audit | | Geographic reach | Markets you need to access | | Technical expertise | Experience with your technology | | Fee structure | Transparency, predictability |