Get Started Free →EU MDR 2017/745 compliance specialist for medical device classification, technical documentation, clinical evidence, and post-market surveillance. Use for Annex VIII classification, GSPR, clinical evaluation, EUDAMED, and UDI.
.claude/skills/borghei-mdr-745-specialist/SKILL.md| Test case | Without → With | Effect | Δ tokens | Δ turns |
|---|---|---|---|---|
| case-02 | ✗→✓ | ▲ Improved | 191% | 0% |
| case-11 | ✗→✓ | ▲ Improved | 378% | 0% |
| case-15 | ✓→✓ | = Same ✓ | 201% | 0% |
| case-03 | ✓→✓ | = Same ✓ | 440% | 0% |
| case-04 | ✓→✓ | = Same ✓ | 833% | 0% |
EU MDR compliance patterns for medical device classification, technical documentation, and clinical evidence.
Before classifying the device or analyzing gaps, confirm these inputs. If any is unknown or vague, ASK — do not assume:
Stop rule: ask only the 2-3 that most change the output. If the user says "just draft it," proceed and list your assumptions at the top of the classification.
Classify device under MDR Annex VIII:
| Factor | Class I | Class IIa | Class IIb | Class III | |--------|---------|-----------|-----------|-----------| | Duration | Any | Short-term | Long-term | Long-term | | Invasiveness | Non-invasive | Body orifice | Surgical | Implantable | | System | Any | Non-critical | Critical organs | CNS/cardiac | | Risk | Lowest | Low-medium | Medium-high | Highest |
| Information Use | Condition Severity | Class | |-----------------|-------------------|-------| | Informs decision | Non-serious | IIa | | Informs decision | Serious | IIb | | Drives/treats | Critical | III |
Example 1: Absorbable Surgical Suture
Example 2: AI Diagnostic Software
Example 3: Cardiac Pacemaker
Prepare technical file per Annex II and III:
ANNEX II TECHNICAL DOCUMENTATION
├── Device description and UDI-DI
├── Label and instructions for use
├── Design and manufacturing info
├── GSPR compliance matrix
├── Benefit-risk analysis
├── Verification and validation
└── Clinical evaluation report| Requirement | Evidence | Status | |-------------|----------|--------| | Safe design (GSPR 1-3) | Risk management file | ☐ | | Chemical properties (GSPR 10.1) | Biocompatibility report | ☐ | | Infection risk (GSPR 10.2) | Sterilization validation | ☐ | | Software requirements (GSPR 17) | IEC 62304 documentation | ☐ | | Labeling (GSPR 23) | Label artwork, IFU | ☐ |
| Class | Route | NB Involvement | |-------|-------|----------------| | I | Annex II self-declaration | None | | Is/Im | Annex II + IX/XI | Sterile/measuring aspects | | IIa | Annex II + IX or XI | Product or QMS | | IIb | Annex IX + X or X + XI | Type exam + production | | III | Annex IX + X | Full QMS + type exam |
Develop clinical evidence strategy per Annex XIV:
| Class | Minimum Evidence | Investigation | |-------|------------------|---------------| | I | Risk-benefit analysis | Not typically required | | IIa | Literature + post-market | May be required | | IIb | Systematic literature review | Often required | | III | Comprehensive clinical data | Required (Article 61) |
CER CONTENTS
├── Executive summary
├── Device scope and intended purpose
├── Clinical background (state of the art)
├── Literature search methodology
├── Data appraisal and analysis
├── Safety and performance conclusions
├── Benefit-risk determination
└── PMCF plan summaryEstablish PMS system per Chapter VII:
| Component | Requirement | Frequency | |-----------|-------------|-----------| | PMS Plan | Article 84 | Maintain current | | PSUR | Class IIa and higher | Per class schedule | | PMCF Plan | Annex XIV Part B | Update with CER | | PMCF Report | Annex XIV Part B | Annual (Class III) | | Vigilance | Articles 87-92 | As events occur |
| Class | Frequency | |-------|-----------| | Class III | Annual | | Class IIb implantable | Annual | | Class IIb | Every 2 years | | Class IIa | When necessary |
| Timeline | Requirement | |----------|-------------| | 2 days | Serious public health threat | | 10 days | Death or serious deterioration | | 15 days | Other serious incidents |
Implement UDI system per Article 27:
| Module | Content | Actor | |--------|---------|-------| | Actor | Company registration | Manufacturer, AR | | UDI/Device | Device and variant data | Manufacturer | | Certificates | NB certificates | Notified Body | | Clinical Investigation | Study registration | Sponsor | | Vigilance | Incident reports | Manufacturer | | Market Surveillance | Authority actions | Competent Authority |
Required elements per Article 13:
references/mdr-classification-guide.md contains:
references/clinical-evidence-requirements.md contains:
references/technical-documentation-templates.md contains:
bash# Quick gap analysis python scripts/mdr_gap_analyzer.py --device "Device Name" --class IIa # JSON output for integration python scripts/mdr_gap_analyzer.py --device "Device Name" --class III --output json # Interactive assessment python scripts/mdr_gap_analyzer.py --interactive
Analyzes device against MDR requirements, identifies compliance gaps, generates prioritized recommendations.
Output includes:
| Factor | Considerations | |--------|----------------| | Designation scope | Covers your device type | | Capacity | Timeline for initial audit | | Geographic reach | Markets you need to access | | Technical expertise | Experience with your technology | | Fee structure | Transparency, predictability |
| Document | Title | Status | Key Impact | |----------|-------|--------|------------| | MDCG 2024-1 | Transition provisions under MDR Art. 120 | Final (2024) | Extended transition deadlines for legacy devices | | MDCG 2024-6 | Clinical evidence needed for medical devices previously CE marked under Directives | Final (2024) | Reduced clinical evidence burden for well-established devices | | MDCG 2023-4 Rev.1 | Notified Body capacity and availability | Revised (2024) | NB capacity monitoring and optimization | | MDCG 2022-18 Rev.1 | Software qualification and classification under MDR | Revised (2024) | Updated software classification algorithm | | MDCG 2020-1 Rev.1 | Clinical evaluation — equivalence | Revised (2024) | Refined equivalence demonstration requirements | | MDCG 2019-16 Rev.1 | Cybersecurity for medical devices | Revised (2024) | Enhanced cybersecurity requirements for connected devices | | MDCG 2019-11 Rev.1 | Qualification and classification of software | Active | Software as medical device classification |
| Device Category | Transition Deadline | Conditions | |----------------|--------------------|-----------| | Class III and implantable | 26 May 2026 | Valid MDD/AIMDD certificate + QMS application to NB by 26 May 2024 | | Class IIb | 31 December 2027 | Valid certificate + QMS application to NB | | Class IIa and Class I (sterile/measuring) | 31 December 2028 | Valid certificate + QMS application to NB | | Class I (up-classified under MDR) | 31 December 2028 | Previously exempt from NB involvement |
Conditions for extended deadlines:
Is the software a medical device?
│
▼
Does the software perform an action on data?
│
Yes─┴─No → NOT a medical device (data storage/communication only)
│
▼
Is the action for the benefit of individual patients?
│
Yes─┴─No → NOT a medical device (population health/admin)
│
▼
Is the action one of: treatment, diagnosis, monitoring, prediction?
│
Yes─┴─No → NOT a medical device (lifestyle/wellness)
│
▼
QUALIFIES AS MEDICAL DEVICE SOFTWARE → Apply classification rules| Decision Factor | Class IIa | Class IIb | Class III | |----------------|-----------|-----------|-----------| | Provides information to inform clinical management | Non-serious conditions | Serious conditions | N/A | | Drives clinical management or diagnoses | N/A | Non-serious conditions | Serious or critical conditions | | Monitors physiological processes | Non-vital parameters | Vital parameters (not immediate danger) | Vital parameters (immediate danger) |
| MDR Requirement | IEC 62304 Mapping | Documentation | |----------------|-------------------|---------------| | GSPR 17.1 (repeatability, reliability) | Software development process | Software development plan | | GSPR 17.2 (state of the art) | Software architecture | Architecture design document | | GSPR 17.3 (minimum IT requirements) | System requirements | IT environment specification | | GSPR 17.4 (foreseeable misuse) | Risk management | Software risk analysis | | Annex II §6.5 (software verification) | Software testing | Test plans and reports | | Annex I §23.4 (labeling for software) | Release documentation | Software release notes |
| AI/ML Capability | MDR Classification Impact | Regulatory Consideration | |-----------------|--------------------------|-------------------------| | AI-assisted detection | Typically Class IIa-IIb (Rule 11) | Must demonstrate clinical performance per intended use | | AI-driven diagnosis | Typically Class IIb-III (Rule 11) | Requires clinical investigation for novel indications | | AI treatment optimization | Typically Class IIb-III (Rule 11 + specific rules) | Benefit-risk analysis must account for AI uncertainty | | Continuously learning AI | Classification per highest risk output | Post-market monitoring must track algorithm evolution |
In addition to standard Annex II requirements, AI/ML devices must document:
| Section | Content | MDCG Reference | |---------|---------|----------------| | Algorithm description | Architecture, training approach, feature engineering | MDCG 2019-11, IMDRF SaMD WG | | Training data | Sources, demographics, size, labeling methodology, quality | MDCG 2020-1 (equivalence) | | Validation methodology | Test dataset independence, performance metrics, subgroup analysis | Annex XIV | | Clinical performance | Sensitivity, specificity, AUC, PPV, NPV per intended population | CER requirements | | Change management | How algorithm updates are validated and deployed | GSPR 17 | | Explainability | How the AI's output can be understood by intended users | GSPR 23 (labeling) |
| EU AI Act Requirement | MDR Equivalent | Combined Approach | |----------------------|----------------|-------------------| | High-risk classification (Annex III, Point 10) | Annex VIII classification rules | Both classifications apply; meet stricter requirement | | Conformity assessment (Art. 43) | Annex IX/X/XI assessment | MDR conformity assessment satisfies AI Act (Art. 120) | | Technical documentation (Annex IV) | Annex II technical documentation | Extend MDR technical file with AI Act-specific elements | | Risk management (Art. 9) | ISO 14971 + GSPR | ISO 14971 satisfies both when AI risks are included | | Data governance (Art. 10) | GSPR 17 + Annex XIV | Add AI-specific data governance to clinical evaluation | | Post-market monitoring (Art. 72) | Chapter VII PMS | Single PMS system covering both AI Act and MDR |
> See also: ../risk-management-specialist/SKILL.md for AI-specific risk management under ISO 14971, and ../fda-consultant-specialist/SKILL.md for FDA's AI/ML SaMD framework and PCCP.
| Module | Status | Mandatory Date | Content | |--------|--------|----------------|---------| | Actor Registration | Operational | Available now | Economic operator registration | | UDI/Device Registration | Operational | 6 months after Eudamed fully functional | Device and UDI-DI data | | Notified Body and Certificates | Operational | Available now | Certificate data upload by NBs | | Clinical Investigations | Operational | Available now | Study registration and reporting | | Vigilance | Partially operational | 24 months after fully functional | Incident reports, FSCAs, trend reports | | Market Surveillance | In development | 18 months after fully functional | CA market surveillance activities |
Key consideration: Until Eudamed is declared fully functional by the European Commission, manufacturers must use existing national systems (e.g., BfArM in Germany, ANSM in France) for vigilance reporting.
| Data Element | Required For | Update Frequency | |-------------|-------------|-----------------| | SRN (Single Registration Number) | All manufacturers | On change | | Authorized representative details | Non-EU manufacturers | On change | | Device identification (UDI-DI) | All devices placed on market | Before first placing on market | | Basic UDI-DI | Device model/family grouping | Before first placing on market | | GMDN code | Device nomenclature | On initial registration | | Risk class | Classification per Annex VIII | On initial registration | | NB certificate reference | Class IIa and above | When certificate issued | | Clinical investigation registration | Interventional studies | Before study start |
| Element | Description | Example | |---------|-------------|---------| | Device identifier | Unique code for device model/version | 04069876543219 (GS1 GTIN) | | Issuing entity | GS1, HIBCC, ICCBBA, or IFA | Selected by manufacturer | | Device model | Specific device configuration | "CardioMonitor X200" | | Device version | Software version (for SaMD) | "v3.1.0" | | Applicable regulations | MDR or IVDR reference | MDR 2017/745 | | Risk class | Per Annex VIII | Class IIa | | Basic UDI-DI | Grouping identifier for device family | 04069876500001 |
| PI Element | When Required | Format | |-----------|---------------|--------| | Lot/batch number | When tracking by lot | Lot: ABC123 | | Serial number | When individual tracking required (Class III, implantable) | SN: 2024-00001 | | Manufacturing date | When relevant to safety | Mfg: 2024-06-15 | | Expiry date | When device has shelf life | Exp: 2026-06-15 | | Software version | SaMD and software-driven devices | SW: v3.1.0 |
| Carrier Type | Format | Where Applied | |-------------|--------|---------------| | AIDC (barcode/2D code) | GS1 DataMatrix, GS1-128, HIBC | Device label, package label | | HRI (human-readable) | Plain text interpretation of AIDC | Adjacent to AIDC on label | | RFID | GS1 EPC/RFID | Optional, in addition to AIDC |
Labeling placement rules:
| Device Class | Submission Deadline | |-------------|-------------------| | Class III and implantable | Before placing on the market | | Class IIb | Before placing on the market | | Class IIa | Before placing on the market | | Class I | Before placing on the market |
> Note: All timelines are contingent on Eudamed being declared fully functional. Until then, manufacturers should pre-register in Eudamed (available modules) and maintain data readiness.
Healthcare organizations manufacturing or deploying medical devices may be classified as essential entities under NIS2:
| NIS2 Requirement | MDR Impact | Action for Manufacturers | |-----------------|-----------|--------------------------| | Art. 21: Cybersecurity risk management | MDCG 2019-16 cybersecurity guidance | Align device cybersecurity with organizational NIS2 compliance | | Art. 23: Incident reporting (24h/72h) | Art. 87-92 vigilance reporting | Unified incident reporting covering both device and infrastructure incidents | | Art. 21(2)(d): Supply chain security | Art. 11 authorized representatives, supply chain | Assess cybersecurity of device component suppliers | | Art. 20: Governance and accountability | Art. 10 manufacturer obligations | Senior management oversight of both NIS2 and MDR compliance |
> See also: ../information-security-manager-iso27001/SKILL.md for ISO 27001 alignment with NIS2 requirements.
eu-ai-act-specialist for AI Act obligationsnis2-directive-specialist for NIS2 compliance| Problem | Possible Cause | Resolution | |---------|---------------|------------| | Device classification unclear -- rules yield different results | Multiple classification rules apply; highest class must be selected per MDR Article 51(7) | Apply all applicable rules from Annex VIII (Rules 1-22); use the implementing rule that gives the highest classification; for software, apply MDCG 2019-11 Rev.1 algorithm; document rationale for each rule considered | | Notified Body rejects technical file for incompleteness | GSPR compliance matrix gaps, missing clinical evaluation, or insufficient risk management documentation | Review GSPR checklist in this skill; ensure Annex II technical file structure is complete; verify CER meets Annex XIV requirements; confirm ISO 14971 risk management file is current and comprehensive | | EUDAMED registration delays blocking market access | Module not operational or manufacturer SRN not obtained | Check current EUDAMED module status; obtain SRN via actor registration module (operational); use national systems for vigilance reporting until Eudamed Vigilance module is fully functional | | Clinical evidence insufficient for Class IIb/III device | Equivalence route rejected by NB or clinical investigation not planned | Reassess equivalence per MDCG 2020-1 Rev.1 (technical, biological, clinical equivalence with access to data); if equivalence fails, plan clinical investigation per Article 61; consider MDCG 2024-6 for reduced evidence burden on well-established devices | | MDR transition deadline approaching with NB application pending | Limited NB capacity (approximately 40 designated EU-wide as of 2025) | Verify transition deadline for your device class (Class III: May 2026, Class IIb: Dec 2027, Class IIa: Dec 2028); ensure QMS application was submitted to NB by applicable deadline; maintain MDD/AIMDD compliance during transition | | UDI labeling rejected by NB or competent authority | UDI-DI/UDI-PI format incorrect, AIDC carrier unreadable, or missing elements | Verify all required UDI elements per Article 27; ensure AIDC format (GS1 DataMatrix preferred) is scannable; include HRI adjacent to barcode; for reusable devices, apply direct marking that survives reprocessing | | AI/ML medical device faces dual regulatory obligations | Device classified under both MDR and EU AI Act as high-risk | Assess both MDR Annex VIII classification and EU AI Act Annex III categorization; MDR conformity assessment may satisfy AI Act per Art. 120; extend technical documentation with AI-specific elements per MDCG 2024-8 |
mdr_gap_analyzer.py, with all requirements addressed or in-progress with documented timelineIn Scope:
Out of Scope:
quality-manager-qms-iso13485 for ISO 13485 QMSrisk-management-specialist for ISO 14971Important Notes:
| Skill | Integration | When to Use | |-------|-------------|-------------| | fda-consultant-specialist | Cross-framework mapping for dual US/EU market; FDA QMSR aligns with ISO 13485 used by MDR | When device requires both FDA clearance/approval and EU MDR CE marking | | quality-manager-qms-iso13485 | ISO 13485 QMS is prerequisite for MDR conformity assessment (Annex IX, XI) | When establishing or auditing QMS for MDR compliance | | risk-management-specialist | ISO 14971 risk management file is core component of MDR technical documentation | When developing risk management file, FMEA, and benefit-risk analysis | | eu-ai-act-specialist | AI medical devices subject to both MDR and EU AI Act; classification and conformity assessment interaction | When AI-enabled medical device requires dual regulatory compliance | | capa-officer | CAPA process supports MDR vigilance obligations and FSCA implementation | When post-market surveillance identifies safety or performance issues requiring corrective action | | infrastructure-compliance-auditor | Cybersecurity validation per MDCG 2019-16 Rev.1 for connected medical devices | When connected device requires cybersecurity documentation for technical file |
Analyzes device against MDR requirements, identifies compliance gaps, and generates prioritized recommendations.
| Flag | Required | Description | |------|----------|-------------| | --device <name> | Yes (unless --interactive) | Device name for gap analysis | | --class <class> | Yes (unless --interactive) | Device classification: I, Is, Im, IIa, IIb, III | | --output <format> | No | Output format: json for machine-readable output; default is human-readable text | | --interactive | No | Launch interactive assessment mode with guided questions |
Analysis Categories: Technical documentation (Annex II), GSPR compliance, clinical evidence (Annex XIV), post-market surveillance (Chapter VII), UDI/EUDAMED, labeling (Article 13), quality management system, risk management, and conformity assessment route.
Output: Requirements checklist with per-item status (Not Started/In Progress/Complete/N/A), gap identification with priority (Critical/High/Medium/Low), critical gap highlighting, completion percentage, and compliance roadmap recommendations.
| Case | Status | Duration (ms) | Turns | Tokens | Tool calls | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Without | With | Δ | Without | With | Δ | Without | With | Δ | Without | With | Δ | ||
case-22 | fail→fail | 20,349 | 20,353 | +0% | 1 | 1 | 0% | 4,092 | 11,669 | +185% | 0 | 0 | — |
case-15 | pass→pass | 20,951 | 16,479 | -21% | 1 | 1 | 0% | 3,428 | 10,331 | +201% | 0 | 0 | — |
case-21 | fail→fail | 29,828 | 37,703 | +26% | 1 | 1 | 0% | 5,118 | 13,808 | +170% | 0 | 0 | — |
case-01 | fail→fail | 35,573 | 33,492 | -6% | 1 | 1 | 0% | 6,207 | 13,459 | +117% | 0 | 0 | — |
case-02 | fail→pass | 21,008 | 14,139 | -33% | 1 | 1 | 0% | 3,410 | 9,928 | +191% | 0 | 0 | — |
case-03 | pass→pass | 9,947 | 11,792 | +19% | 1 | 1 | 0% | 1,791 | 9,679 | +440% | 0 | 0 | — |
case-04 | pass→pass | 5,738 | 6,717 | +17% | 1 | 1 | 0% | 944 | 8,810 | +833% | 0 | 0 | — |
case-05 | pass→pass | 5,921 | 7,493 | +27% | 1 | 1 | 0% | 971 | 8,869 | +813% | 0 | 0 | — |
case-06 | pass→pass | 16,143 | 19,378 | +20% | 1 | 1 | 0% | 2,794 | 10,713 | +283% | 0 | 0 | — |
case-07 | pass→pass | 16,903 | 14,646 | -13% | 1 | 1 | 0% | 2,721 | 10,054 | +269% | 0 | 0 | — |
case-08 | fail→fail | 8,334 | 10,573 | +27% | 1 | 1 | 0% | 1,525 | 9,568 | +527% | 0 | 0 | — |
case-09 | pass→pass | 7,517 | 8,257 | +10% | 1 | 1 | 0% | 1,363 | 9,039 | +563% | 0 | 0 | — |
case-10 | fail→fail | 9,821 | 10,848 | +10% | 1 | 1 | 0% | 1,776 | 9,597 | +440% | 0 | 0 | — |
case-11 | fail→pass | 10,149 | 7,079 | -30% | 1 | 1 | 0% | 1,845 | 8,822 | +378% | 0 | 0 | — |
case-12 | pass→pass | 9,800 | 10,449 | +7% | 1 | 1 | 0% | 1,642 | 9,447 | +475% | 0 | 0 | — |
case-13 | fail→fail | 8,630 | 11,477 | +33% | 1 | 1 | 0% | 1,617 | 9,633 | +496% | 0 | 0 | — |
case-14 | pass→pass | 10,692 | 10,787 | +1% | 1 | 1 | 0% | 1,832 | 9,491 | +418% | 0 | 0 | — |
case-16 | fail→fail | 6,961 | 3,297 | -53% | 1 | 1 | 0% | 1,051 | 8,178 | +678% | 0 | 0 | — |
case-17 | pass→pass | 13,580 | 9,928 | -27% | 1 | 1 | 0% | 2,374 | 9,198 | +287% | 0 | 0 | — |
case-18 | pass→pass | 16,467 | 15,369 | -7% | 1 | 1 | 0% | 2,547 | 9,965 | +291% | 0 | 0 | — |
case-19 | pass→pass | 15,548 | 13,670 | -12% | 1 | 1 | 0% | 2,419 | 9,796 | +305% | 0 | 0 | — |
case-20 | fail→fail | 22,630 | 33,467 | +48% | 1 | 1 | 0% | 4,045 | 13,371 | +231% | 0 | 0 | — |
DecimalAI ran this skill against gemini-3.6-flash twice over the same eval suite — once with the skill loaded and once without — and compared the two runs case by case. 22 cases were attempted. The headline lift of +9 percentage points is the difference between those two pass rates over the 22 comparable cases.